8. Error Messages
8
Display /
Problem
Display Meaning Possible cause and
fault remedy
Measured
temperature too
high
Displays «H» when
measured temperature
higher than 42.2 °C
(108.0 °F).
Measured
temperature too low
Displays «L» when
measured temperature
lower than 34.0 °C
(93.2 °F).
Ambient
temperature too
high
Displays «H» in
conjunction with the
«» when ambient
temperature is higher
than 40.0 °C or 104.0 °F.
Ambient
temperature too low
Display «L» in
conjunction with the
«» when ambient
temperature is lower
than 16.0 °C or 60.8 °F.
Error function
display
When system has
malfunction.
Blank display Please check if the
battery has been loaded
correctly. Also check
polarity (<+> and <->) of
battery.
Dead battery
indication
If the steady battery
icon is the only symbol
shown on the display,
the battery should be
replaced immediately.